Fault
Cause
Solution
Instruments
do not hold
in place
Permitted maximum speed
for the instrument in use has
been exceeded
Instrument shaft worn
Instrument shaft bent
Instrument not correctly
inserted into the clamping
device
Cream, ointment, or similar
on the instrument shaft
Working pressure too high on
the instrument
Pulling motions during work
Only work up to the
maximum permitted speed of
the instruments used
Check the instrument shaft
and use a new instrument if
necessary
The instrument must be
replaced
Push the instrument
completely into the clamping
device as far as the stop
Clean the instrument.
Work with small instruments
until larger instruments start
to clamp securely again.
During operation, always
make sure that the
handpiece tip and the
instruments are clean.
Reduce pressure
Avoid pulling motions
Suction turbine
switches to
suction level 3
during
operation
Dust bag, turbine protection filter
and filter cassette are full, device
switches automatically to suction
level 3 in order to avoid damage
due to rising temperature.
Change all filters and the dust
bag and make sure that the
extracted air can escape easily
from the back of the device.
Handpiece
vibrates;
handpiece is
loud
with
instruments
inserted
Maximum permitted speed
for instrument in use is
exceeded,
Instrument is defective (bent,
shaft worn, etc.)
Only work up to the
maximum permitted speed
of the instrument used
Use new instruments.
Push-button on
handpiece jams
or is difficult to
operate
The handpiece or the handpiece
button is dirty
Clean the push-button
If one of these faults still persists despite having followed this information, please send
the unit to the customer service team.
In addition, please get in touch with the manufacturer for each fault / each malfunction
in cases of doubt.