BaehrTec A2000 operating instructions
Page 50/58
English – V101 - effective 10/13
Apply a little oil with a cotton swab to the button pusher and reinsert it into the button
housing.
Fig. 29
Checker that the functioning is back to normal by pushing the button pusher with your
finger.
Fig. 30
5.5.3
Changing the dampers for the handpiece case
If you notice that the handpiece motor no longer sits comfortably in the handpiece
case, this means that the dampers for the handpiece case are worn.
To change the dampers for the handpiece case, please proceed as follows:
Switch off your device at the main switch and disconnect the device from the mains.
Unscrew the handpiece case from the handpiece cap.
Fig. 31
Fig. 32