XM/SM/XE MAGNETIC FLOWMETER
CALIBRATION
WBPEEUI110501A0
E - 7
D-to-A Adjust
The
D-TO-A ADJUST
selection is only present when you are in the
ANA-
LOG
communication mode, specified in the configuration procedure. This
selection allows you to adjust the four to 20 milliampere output of the field
device.
Empty pipe detector calibration is complete.
Key
Display
Comments
ENTER
CONFIG. ID TAG
READY
Key
Display
Comments
Select
D-TO-A ADJUST
.
Use the arrow keys to adjust the 4 mA signal.
NOTE:
When increasing or decreasing the mA signal, the
increments of change increase with successive depressions
until the maximum level of change is reached. By chang-
ing direction you will return to the smallest increment of
change. This adjustment technique speeds up the adjust-
ment process without affecting fine adjustment.
Use the arrow keys to adjust the 20 mA signal.
CALI-
BRATE
OUTPUT WILL BE
AFFECTED!
PROCEED?
NO
→
YES
ENTER
CALIBRATION
→
D-TO-A ADJUST
EMPTY PIPE DET
ENTER
ADJUST TO 4 mA
THEN HIT ENTER
ENTER
ADJUST TO 20 mA
THEN HIT ENTER
ENTER
CONFIG. ID TAG
READY