Table 7: Modbus register map
Function
Parameter
Description
Range/state
Addr
Data type
Read
only
Settings
Adjust
PPM level offset adjust
-25.00 ~ +225.00
5210
Double/float
Moisture reading
average filter size
10 ~ 200 samples
5230
Integer
Clock
Hour
0~23
5410
Integer
Minutes
0~59
5420
Integer
Month
1~12
5430
Integer
Date
1~28/29/30/31
5440
Integer
Year
2000~2099
5450
Integer
External pressure
Constant
0 ~ 3500.00 kPa
5510
Double/float
Pressure zero
calibration, mA
0~22 mA
5520
Double/float
Pressure span
calibration, mA
0~22 mA
5525
Double/float
Pressure zero
calibration, kPa
0~3500 kPa
5530
Double/float
Pressure span
calibration, kPa
0~3500 kPa
5535
Double/float
Pressure source
Constant value = 0,
live sensor = 1
5540
Integer
Device ID
Aurora H2O
serial
number
------
------
8100
8 character
bytes
•
Laser serial number
------
------
8200
8 character
120 bytes
•
Calibration date
Month
1-12
8310
Integer
•
Date
Depends on month
8320
Integer
•
Year
2000~2100
8330
Integer
•
System up time
MSDate
Uptime, in days
8400
Double/float
•
Measurements Dew point
Dew point °C
------
9110
Double/float
•
Dew point °F
------
9120
Double/float
Equivalent
dew point °C
------
9130
Double/float
•
Equivalent
dew point °F
------
9140
Double/float
•
Temperature
Sample
temperature °C
------
9210
Double/float
•
Sample
temperature °F
------
9220
Double/float
•
External Pressure
kPa
------
9510
Double/float
•
MPa
------
9512
Double/float
•
PSIa
------
9520
Double/float
•
78
Summary of Contents for Panametrics Aurora H2O
Page 2: ...ii...
Page 4: ...no content intended for this page proceed to next page iv...
Page 20: ...no content intended for this page 10...
Page 32: ...no content intended for this page 22...
Page 71: ...Figure 74 Exporting a simplified image Figure 75 Selecting enhanced metafile 61...