4.3.4 Adjusting the gas offset
The input information for
gas molecular weight is
not currently used for any
moisture calculations and
is reserved for future use.
• lbs/MMSCF is calculated using IGT research Bulletin #8
and ASTM D-1142-95 referenced at 60°F, 1 ATM.
• mg/cm
3
is based on ideal gas law derivation referenced
at 15°C, 1.01325 kPa.
To adjust the gas molecular weight offset, from the settings
menu select gas and press enter. From the gas data menu
select mol. weight and press enter. The following screen
appears.
Use the left and right
arrow keys to select each
digit to be changed. Use
the up and down arrow
keys to change the value.
When finished, press enter.
The screen returns to the
settings menu.
4.4 Clock settings
The clock settings are for
informational purposes.
They are used to keep
track of the test analyzer
start time and the laser
operational time.
To reset the clock, from
the settings menu select
clock and press enter. The
following screen appears.
4.4.1 Resetting the hour
To reset the hour, from
the clock menu select
hour and press enter. The
following screen appears.
Use the left and right
arrow keys to select each
digit to be changed. Use
the up and down arrow
keys to change the value.
When finished, press enter.
The screen returns to the
clock menu.
4.4.2 Resetting the minutes
To reset the minutes, from
the clock menu select
minutes and press enter.
The following screen
appears.
Use the left and right
arrow keys to select each
digit to be changed. Use
the up and down arrow
keys to change the value.
When finished, press enter.
The screen returns to the
clock menu.
41
Summary of Contents for Panametrics Aurora H2O
Page 2: ...ii...
Page 4: ...no content intended for this page proceed to next page iv...
Page 20: ...no content intended for this page 10...
Page 32: ...no content intended for this page 22...
Page 71: ...Figure 74 Exporting a simplified image Figure 75 Selecting enhanced metafile 61...