24
To configure the XDP output, navigate to the
cal>1:cal>setup>analog output
menu and proceed as follows:
1. At the
analog output
menu option press [ENTER].
2. Use the [
] and [
] keys to highlight the
Aout type
option and press [ENTER].
3. Use the [
] and [
] keys to highlight the menu option
for the desired output type and press [ENTER].
4. Use the [
] and [
] keys to highlight the
zero setpoint
option and press [ENTER].
5. Using the four arrow keys, program the mA/pcnt
measurement value that corresponds to the low end
of the chosen analog output range. Press [ENTER].
6. Use the [
] and [
] keys to highlight the
span setpoint
option and press [ENTER].
7. Using the four arrow keys, program the ppm/pcnt
measurement value that corresponds to the upper end
of the chosen analog output range. Press [ENTER].
4.6.2a Error handler options
At this point, the programming of the XDP analog output
differs from that of the system analog output described in
the previous section. The difference lies in the programming
of error handlers for the following six fault conditions:
• Input mA over range
• Input mA under range
• Zero drift/cal > max drift/cal
• Zero drift/total > max drift/total
• Span drift/cal > max drift/cal
• Span drift/total > max drift/total
NOTE:
See
"Chapter 5. Calibration" on page 27
, for
instruction on viewing the current values for last four
parameters listed above.
4.6.2b Configuring the error handlers
To set up the error handlers, continue the programming
as follows:
1. At the
error handlers
menu option press [ENTER].
2. Use the [
] and [
] keys to highlight the desired
fault condition
error handler (see list on previous page)
and press [ENTER].
3. Use the [
] and [
] keys to highlight either the
enable
or the
disable
option and press [ENTER].
4. If you
disabled
this error handler, skip to step 7. If you
enabled
this error handler continue with the next step.
5. Use the [
] and [
] keys to highlight the manner in which
you want the XDP to respond to a fault in this parameter:
•
Force high
- displays a value equal to the high point
of the selected analog output range (e.g. 20 mA for a
4-20 mA range).
•
Force low
- displays a value equal to the low point
of the selected analog output range (e.g. 4 mA for
a 4-20 mA range).
•
Force to value
- displays a preprogrammed value.
•
No effect
- displays the current live value
•
Hold last value
- displays the last valid measurement
prior to the fault condition.
After making your choice, press [ENTER].
6. If you selected
force high, force low, no effect
, or
hold
last value
, go to step 7. If you selected force to value,
proceed as follows:
a. Using the four arrow keys, program the ppm/pcnt
measurement value that should be displayed if there
is a fault condition in this parameter. Press [ENTER].
b. Press [ESC] twice.
7. Return to step 2 to program additional error handlers,
until all of the error handlers have been configured.
4.6.2c Testing and trimming the analog output
Having set the basic parameters for the analog output, the
output should now be tested and trimmed by continuing
as follows:
1. Use the [
] and [
] keys to highlight the test option
and press [ENTER].
2. Enter a test percentage (0-100%) based on the full scale
value for the chosen analog output type (e.g. a 50%
test value for the 4-20 mA range would be 12 mA). Press
[ENTER] to leave the edit box and press [ENTER] again at
either the
test, zero trim
, or
span trim
menu choice to
send the test value to the analog output device.
3. Repeat step 2 for as many test percentages as desired,
then press [ENTER].
IMPORTANT
Before proceeding, make sure the output percentage to be
trimmed has been entered in the test menu.
4. Use the [
] and [
] keys to highlight the
zero trim
option
and press [ENTER].
5. Use the [
] and [
] keys to trim the zero point until
your output device reads the exact value for the low
end of the chosen analog output range (e.g. 4 mA for
a 4-20 mA range). Press [ENTER].
6. Use the [
] and [
] keys to highlight the
span trim
option
and press [ENTER].
7. Use the [
] and [
] keys to trim the span point until
your output device reads the exact value for the high
end of the chosen analog output range (e.g. 20 mA for
a 4-20 mA range). Press [ENTER].
Summary of Contents for Panametrics XDP
Page 1: ...Gas analysis XDP User s manual 910 204 Rev E...
Page 2: ...ii...
Page 4: ...no content intended for this page proceed to next page iv...
Page 8: ...viii no content intended for this page...
Page 12: ...no content intended for this page proceed to next page 2...
Page 28: ...no content intended for this page proceed to next page 18...
Page 36: ...26 no content intended for this page proceed to next page...
Page 42: ...32 no content intended for this page proceed to next page...
Page 58: ...48 no content intended for this page proceed to next page...