Bugbox Plus
UM-17 R4.0
Page 34 of 47
Bugbox Plus User Manual
Cleaning procedure – deep clean
To deep clean the workstation;
Preparing the workstation
Remove all cells/ samples to an alternative storage facility.
Switch the workstation off at the mains and remove the plug from the mains.
Remove the Ezee Sleeves
(see section 7.2.2, for more details).
Remove any other equipment installed in the workstation chamber.
Cleaning the workstation chamber
The items inside the workstation chamber that require cleaning are;
o
The floor tray.
o
The ceiling panel.
o
The shelf panel.
o
The left hand wall.
o
The right hand wall.
o
The interlock inner door.
o
The glove port covers.
o
If fitted, the inside of the SPES.
For all workstation components, wipe with a paper towel soaked in an appropriate cleaning
agent. Take care not to get cleaning agents on the fans (below the shelf panel) and if fitted,
the plug socket.
Wipe dry using paper towels.
Cleaning the interlock chamber;
The items inside the interlock chamber that require cleaning are;
o
The floor.
o
The interlock outer door.
o
The Petri dish holder.
o
The inside wall.
o
The interlock inner door.
For all interlock components, wipe with a paper towel soaked in an appropriate cleaning
agent.
Wipe dry using paper towels.
Reinstalling workstation components
Reinstall any equipment removed from the workstation chamber.
Reinstall the Ezee Sleeves (see section 7.2.2, for more details).