Bugbox Plus
UM-17 R4.0
Page 40 of 47
Bugbox Plus User Manual
Service requirements
To maintain the best performance from your Bugbox Plus workstation, it must be serviced at regular
intervals. Figure 32 lists the servicing requirements, intervals and persons capable of performing the
service;
Figure 32: Bugbox Plus servicing requirements
Action
Frequency
By
Clean workstation
After each use
End User
Fill humidifier tank
Weekly
End User
Deep clean workstation
3-6 months depending on usage End User
Replace detox sachet
Annually
End User
Replace catalyst sachet
Annually
End User
Annual service
Annually
Qualified service engineer
Biennial service
Biennially (2 yearly)
Qualified service engineer
To arrange an annual or biennial service, contact your local distributor.
Note: The biennial service includes an annual service.
Note: The annual service includes the replacement of the detox and the catalyst sachet.
Service contracts are available for all Ruskinn Technology Limited workstations. Please contact your
local distributor for more information.
Spare parts and accessories
A range of spare parts and accessories are available for your Bugbox Plus workstation.
Note: Only Ruskinn Technology Limited spare parts should be used. The use of unapproved spare parts
will invalidate the warranty of your workstation and may cause damage to your workstation.
Overview
To order spare parts and accessories, please contact your local distributor for the latest pricing and
availability.