Optional accessories
6.8
Your Invivo
2
Plus workstation may be fitted with optional accessories to provide added functionality.
The available optional accessories for Invivo
2
Plus workstations are;
•
Cable gland.
•
Universal cable gland.
•
Gas sample port.
•
Ultrasonic humidity control.
•
Vacuum port.
•
HEPA protection option (see section 6.2).
Note that these optional accessories are not available as aftermarket upgrades; they must be fitted
during the manufacture of the workstation.
Cable gland
6.8.1
The cable gland is used to allow cables to enter the workstation chamber without affecting the
internal environment of the workstation chamber. The cable gland is suitable for cables of diameter
3.5mm to 7mm. The cable gland is located on the left hand side of the workstation
Figure 31 shows the cable gland;
Figure 31: Cable gland
To use the cable gland;
•
Turn the grey collar anti-clockwise (counter-clockwise) to loosen the cable gland.
•
Remove the red plug.
•
Push the cable through the cable gland.
•
Tighten the grey collar by turning clockwise until tight. Do not over tighten the cable gland.
Universal cable gland
6.8.2
The Universal cable gland is used to allow cables to enter the workstation chamber without affecting
the internal environment of the workstation chamber. The Universal cable gland provides a diameter
of 50mm for passing through larger cables and connectors. The Universal cable gland is located on
the rear of the workstation.