Invivo
2
Plus
UM-14 Version R3.0
Page 44
Invivo
2
Plus User Manual
Replacing an Ezee Sleeve
7.4.2
To remove an Ezee Sleeve;
•
Ensure that the glove port covers are closed.
•
Remove the O-rings that hold the Ezee Sleeve to the glove port. Note that the O-rings will be
tight. Figure 36 shows the O-ring removal;
Figure 36: O-ring removal
•
Remove the Ezee Sleeve
To install an Ezee Sleeve;
•
Place the new Ezee Sleeve onto the glove port, ensuring that the O-ring grooves are
completely covered and that the Ezee Sleeve is not kinked or overlapping.
•
Reinstall the O-rings over the Ezee Sleeve. Note that the O-rings will be a tight fit. This is
intentional, to provide a gas tight seal. Note that the O-rings must completely cover the Ezee
Sleeve to provide a gas tight seal.
Replacing detox sachets
7.4.3
Detox sachets are supplied with the Invivo
2
Plus workstation. The detox sachets adsorb volatile
organic compounds, improving the air quality within the workstation chamber. Detox sachets need
to be replaced annually. See section 7.4.2 for part details. There are 2 methods for replacing the
detox sachets;
•
Via the interlock, to maintain the atmospheric condition in the workstation chamber.
•
With the front screen removed, as part of a deep clean.