Invivo
2
Plus
UM-14 Version R3.0
Page 46
Invivo
2
Plus User Manual
To reinstall the detox sachets with the front screen removed;
•
Remove the new detox sachets from their packaging.
•
Whilst lifting the front of the floor tray, slide the detox sachets into the detox sachet
holders.
•
Lower the front of the floor tray back to its original position.
•
Reinstall any equipment previously installed in the workstation.
•
Reinstall the Ezee Sleeves (see section 7.4.2 for more details).
•
Cells can be reintroduced to the workstation from the alternative storage facility.
Note:
It is recommended to perform a deep clean (see section 7.3 for more details) when replacing
the detox sachets with the front screen removed.
Installing/ replacing a catalyst sachet
7.4.4
Catalyst sachets are available as an accessory for the Invivo
2
Plus workstation. See section 7.6 for
ordering details. The catalyst sachets are required for anaerobic operation of the workstation. There
are 2 methods for replacing the detox sachets;
•
Via the interlock, to maintain the atmospheric condition in the workstation chamber.
•
With the front screen removed, as part of a deep clean.
To install a catalyst sachet via the interlock;
•
Remove the catalyst sachet from its packaging.
•
Open the interlock outer door (see section 6.1.2 for more details).
•
Place the catalyst sachet in the interlock chamber.
•
Close the interlock outer door (see section 6.1.2, for more details).
•
Start the interlock purge cycle by pressing the interlock cycle start button (item 2 in Figure
13: Interlock outer door control panel).
•
Enter the workstation chamber via the Ezee Sleeves (see section 6.2 for more details).
•
Move any samples/ equipment onto the rear shelf. Do not exceed the weight limit for the
rear shelf.
•
Lift the front of the floor tray to expose the detox sachets and sachet holders.
•
Whilst holding the floor tray up, slide one detox sachet forward and remove from sachet
holder.
•
When the interlock purge cycle has completed, open the interlock inner door (see section
6.1.3 for more details).
•
Place the detox sachet in the interlock chamber and bring the catalyst sachet into the
workstation chamber.
•
Close the interlock inner door (see section 6.1.3, for more details).
•
Whilst lifting the front of the floor tray, slide the catalyst sachet into the detox sachet
holder.
•
Lower the front of the floor tray back to its original position.
•
Move any samples/ equipment back to the floor tray.
•
Exit the workstation chamber via the Ezee Sleeves (see section 6.2 for more details).
•
Open the interlock outer door (see section 6.1.2 for more details).