14
english
6.3
Calibration procedure overview
6.3.1 Teach-in
The factory set null point and end point is replaced by a
new null point and end point. The null point and end point
can be set separately, the output gradient changes.
The detailed procedure for teach-in is described
Steps
►
Move magnet to the new null position.
►
Read new null point by activating the programming
inputs.
⇒
The current end point remains the same.
New null point
Before
After
Fig. 6-2: Reading new null point
►
Move magnet to the new end position.
►
Read new end point by activating the programming
inputs.
⇒
The current null point remains the same.
6
Calibration procedure (continued)
New end point
Before
After
Fig. 6-3: Reading new end point
Only with BTL7-C/E…:
►
The gradient of the current output can be inverted by
activating the programming inputs.
⇒
For example, a rising output gradient is changed to
a falling gradient. The voltage outputs are not
inverted.
BTL7-A/C/E/G5 _ _ -M _ _ _ _ -H/W(8)-S32/S115/_ _ _
Micropulse Transducer - Rod Style
Summary of Contents for BTL7 Series
Page 1: ...Betriebsanleitung BTL7 A C E G5_ _ M_ _ _ _ H W 8 S32 S115 _ _ _ S 3 2 deutsch ...
Page 2: ...www balluff com ...
Page 28: ...User s Guide BTL7 A C E G5_ _ M_ _ _ _ H W 8 S32 S115 _ _ _ S 3 2 english ...
Page 29: ...www balluff com ...
Page 55: ...Manual de instrucciones BTL7 A C E G5_ _ M_ _ _ _ H W 8 S32 S115 _ _ _ S 3 2 español ...
Page 56: ...www balluff com ...
Page 82: ...Notice d utilisation BTL7 A C E G5_ _ M_ _ _ _ H W 8 S32 S115 _ _ _ S 3 2 français ...
Page 83: ...www balluff com ...
Page 109: ...Manuale d uso BTL7 A C E G5_ _ M_ _ _ _ H W 8 S32 S115 _ _ _ S 3 2 italiano ...
Page 110: ...www balluff com ...