Ultrasonic flow metering

BAMOFLONIC

24-04-2014    

776 M1 01 C

INSTRUCTIONS MANUAL

BAMOFLONIC

BAMOFLONIC

Ultrasonic flow metering / dosing device 

Summary of Contents for BAMOFLONIC

Page 1: ...MES 776 01 1 Ultrasonic flow metering BAMOFLONIC 24 04 2014 776 M1 01 C INSTRUCTIONS MANUAL BAMOFLONIC BAMOFLONIC Ultrasonic flow metering dosing device ...

Page 2: ... designed according to the current industry standard EN 61010 corresponds to VDE 0411 Safety specifications for electrical measurement control and laboratory devices The manufacturer cannot be held liable for any damage from inappropriate or unintended use Conversions and or changes to the flowmeter may only be made according to these operating instructions Personnel for installation commissioning...

Page 3: ...lt settings 11 3 4 General Information 11 4 Exchange of measuring device 12 5 Technical specifications 12 5 1 Dimensions and weight 12 5 2 Specifications 14 TABLE OF FIGURES Fig 1 Principle of ultrasonic flow measuring 4 Fig 2 Installation position of BAMOFLONIC 5 Fig 3 Mounting examples of BAMOFLONIC 6 Fig 4 Mounting possibilities of BAMOFLONIC 6 Fig 5 Mounting of BAMOFLONIC 6 Fig 6 Mounting with...

Page 4: ...mounts Chemical resistant Integrated display with keypad 1 2 Measuring principle It usually takes more energy to swim against the flow than with the flow The ultrasonic flow measurement is based on the phase difference approach Two sensors that are located opposite from each other alternatively transmitting and receiving ultrasonic signals With medium at standstill both sensors receive the transmi...

Page 5: ...connection For best measuring performance the BAMOFLONIC should be mounted vertically into the pipe It is not recommended to install the flowmeter after a dosing valve The flowmeter can run empty To avoid getting bubbles in the liquid the BAMOFLONIC should be installed on the pressure side of the pump i IMPORTANT i IMPORTANT NOTICE 2 Assembly and installation 2 1 Installation instructions The hous...

Page 6: ...ations or movements Use the clamps as seen in the figure below at the inlet and outlet connections of BAMOFLONIC Fig 3 Mounting examples of BAMOFLONIC Fig 4 Mounting possibilities of BAMOFLONIC The BAMOFLONIC must be installed without mechanical force on the existing pipe system otherwise the flowmeter may be damaged Pay attention to the axial channel offset of 5mm when mounting the flowmeter For ...

Page 7: ...in the operating limits stipulated on the product label and the operating manual data sheet Use outside these conditions lead to overloads which cause permanent damage 3 Commissioning Note While commissioning run the Basic Trim FlowSoft Medium with filled pipe Repeat this action until amplifier stage and receiving amplitude reach a steady value 3 1 Operation If the BAMOFLONIC is used as volume flo...

Page 8: ...ies to change values a password will be required Password protection is used to ensure just authorized personal is able to change values or configu rations The default password for the BAMOFLONIC is 41414 The password can be changed with FlowSoft The user level will remain active for 5 minutes after the last press on any button Note Not all functions of BAMOFLONIC can be changed in the user menu d...

Page 9: ...diagnostic function which optimizes all important parameters This process lasts approximately 1 minute To make sure the basic trim is correctly done the flowmeter has to be filled with liquid without a flow When there is an error detected while performing the basic trim function the display shows Error After successfully finishing the basic trim the display shows Done Creeping suppression The cree...

Page 10: ...he control Example 2 0 ml Pulse This means a pulse is emitted every 2 0 ml Setting range 0 1 3000 0 ml Pulse in 0 1 ml Pulse steps Default setting 1 0 ml Pulse Dosing By choosing the dosing function via the user display in the manual mode dosing is configurable The Volume Dosing Batch and the Dosing Time are freely adjustable When the Dosing Time is set to zero the timer control is inactive A dosa...

Page 11: ... by pre selecting the meter pulses Dosing starts when the relevant pushbutton of the dosing equipment is pushed The control will now open the dosing valve From now on BAMOFLONIC will send a voltage pulse to the control for each volume unit that has flown through e g per 1ml When the pre selected pulse quantity is reached the control closes the dosing valve In this case output 1 is used to send out...

Page 12: ...must be removed before returning This is particularly important if the medium to be measured is health hazardous Devices judged to be insufficiently cleand will be returned to sender No inspection of device will be done until proper cleaning is completed by user Attach a detailed report describing the failure the application and the physical chemical properties of the medium parameters and the dec...

Page 13: ...1864 3 147 0 84 0 84 5 71 1 15 0 5 0 347 DN20 DIN 11864 3 160 0 84 0 94 2 77 6 15 0 5 0 429 DN25 1 DIN 11864 3 168 0 84 0 98 5 77 6 15 0 5 0 469 DN Connection A mm B mm C mm D mm E mm F mm G Weight g DN10 Tube 160 0 84 0 83 0 70 5 15 0 5 0 12 0 332 DN15 Tube 178 0 84 0 84 5 71 1 15 0 5 0 18 0 344 DN20 Tube 197 0 84 0 94 2 77 6 15 0 5 0 24 0 414 ...

Page 14: ...f the EC directives and has the CE label Digital output Switch over to high resistance when the supply is smaller than 18V When overload or short circuit is detected the digital outputs are switched off after 100µs for a period of 2s When time is up the outputs get applied again Empty pipe output Empty pipe Device filled NPN opener High resistant 0 V NPN shutter 0 V High resistant PNP opener High ...

Page 15: ...llowing graphic min Range is used for 0 and max Range is used for 100 Current mA Current output 0 20 mA For the following graphic min Range is used for 0 and max Range is used for 100 Current mA Value 100 max Range Value Empty pipe Smaller 1 2 Between 0 min Range Between 0 and 100 100 max range Between100 and 103 Bigger 103 Current mA 3 5 3 8 Linear interpolation from 3 8 to 4mA 4 Linear interpola...

Page 16: ...inside the housing Step response time after a significant temperature change Filter Off Influence of the environment temperature Some examples Liquid T x 0 7 environ T x 0 3 measured T 40 C x 0 7 20 C x 0 3 34 C 40 C x 0 7 30 C x 0 3 37 C 40 C x 0 7 40 C x 0 3 40 C 60 C x 0 7 20 C x 0 3 48 C In case of an error In case of a short circuit of the thermocouple 50 C are shown In case of a cable brake ...

Page 17: ...95 7 A Legal directives According to legal obligations and for our staff safety and workshop protection it is necessary to send with the equipment this certificate to declare the decontaminating procedures have been applied before shipment Your responsibility will apply in case of non respect of proper operations 2 2 2 2 3 2 4 3 6 57 7 7 ...