LEVEL INDICATORS
LEVEL INDICATORS
MAGTOP 801, 803, 806
19-03-2014
560 M0 06 C
MES
560-06
/1
WARNING
Level indicator installation and using must be done outside from any magnetic field zone. Use non magnetic connection (Brass, Stainless
Steel or plastic). Any iron device must be installed at 10 cm minimum distance fare. Ensure chemical compliance between liquid and
level indicator. To get easy mounting and dismounting, a valve can be placed between tank and level indicator flange. Those valves may
be used, also, for liquid regulation to avoid strong float up movement (which may damage it). Inlet tank pressure, then in level indicator,
must not be over specified in technical features (6 bar as a maximum).
INTRODUCTION
Those level indicators allow instantaneous liquid level reading, even they are opaque.They are installed on tank side, by flanges.
Measurement on communicating vessels principle. The Float, equipped with magnet, moves bi-colours flaps which keep their positions
until the float is coming back. Their colours (red and yellow on black back side) ensure instantaneous and easy reading. Equipped with
graduated rule (in option) reading accuracy will be higher.
INSTRUCTIONS MANUAL
Level indicators
MAGTOP
801, 803, 806