Operation
3.1
Operating elements
The unit is operated from the front:
1
2
3
4
5
1 LCD display
2 Arrow keys [
], [
], [
], [
]
3 Key [SET]
4 Key [START/STOP]
5 Status LED
3.1.1 Turning the ultrasonic bath on/o
ff
The ultrasonic bath is turned on/o
ff
using the mains switch on the rear side. After turning
on, the LCD display must light up. Initialisation occurs automatically.
=====================
BANDELIN electronic
* S O N O C O O L *
=====================
- initialising -
Next, the ultrasonic bath will be on stand-by mode and will be ready for operation.
The status LED (5) will be lit up, the most recently-set values for runtime, temperature
and ultrasonic power will be displayed.
(Values when activated for the
fi
rst time): target temperature 17 °C, target runtime
00:10:00 hh:mm:ss, target power 100 %).
The last-selected parameters ( runtime, target temperature, target power) will be saved
and accessed the next time the device is turned on.