4.3.1
Mounting and dismounting of the probes
Probes
• are screwed onto the standard or booster horns,
• transmit ultrasonic power into the medium to be sonicated,
• are made of high-strength titanium alloy.
Caution ! Probes are sensitive to shock.
Before mounting the probes, the HF generator must be turned off and the ultrasonic transducer
must be disconnected from the HF generator.
The mounting surfaces
/2
must be thoroughly cleaned so
that the amplitude and/or power can be transmitted to the
medium.
If the mounting surfaces are not cleaned, the probe,
standard horn or booster horn could be destroyed as a
result.
The tool required for mounting/ dismounting is included in the scope of supply.
/2
The mounting surfaces are the contact surfaces between the individual accessory parts, e.g. between the standard or booster horn and
the probe.