UW 50
→
HD 4050
UW 100
→
HD 4100
UW 200
→
HD 4200
UW 400
→
HD 4400
16 / 46
51376-010 GB/2020-10
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com
3.1.1
Turning the homogeniser on/o
ff
The homogeniser is turned on with the power switch (front side, bottom). After turning on, the LC
display must light up. Initialisation occurs automatically. The ultrasonic homogeniser will display
the manufacturer's name, type designation, and the last-connected probe. The type designation
depends on the ultrasonic converter that is connected
BANDELIN electronic
HD 4050
Converter: UW
50
Probe: TS
106
o.k. −
►
[start]
If the probe type displayed does not conform with the probe that is mounted, the correct probe
type will have to be set using the arrow keys
or
, before the next steps. By pressing the
"START/STOP" key, the menu switches into stand-by mode (idle mode) and the ultrasonic
homogeniser is ready for operation. The parameters used during the last operation will be set
again and displayed. If the ultrasonic converter was exchanged in the meantime, the earlier
settings will be replaced by standard values and must be re-edited. The temperature display is
only visible when a temperature sensor is connected.
Prog 3* Ampl
10%
Time set
00:01:00
elapsed
00:00:00
Pulsation off
Energy 0.000
kJ
The mains switch can also be used for powering o
ff
.
The powering-o
ff
process may take a few seconds while the current data is saved internally. The
monitor is turned o
ff
during this process. When powered-on again, the data becomes available
once again.
3.1.2 Explanation of the display
fi
elds
Prog 3* Ampl
10%
Time set
00:01:00
00:00:04
Pulsation off
Energy 0.000
kJ
Preset value
Amplitude (Am) in %
or power (Pw) in W
Preset value
Sonication time
Hours:Minutes:Seconds
(Display omitted during continuous
operation)
Energy applied to the medium
Preset value
Pulsation time OFF
(speci
fi
cation omitted with continuous
operation)
Time elapsed
Hours:Minutes:Seconds
Program No.
with modi
fi
ed programs,
a "*" appears behind the No.
3.2
Setting the operating parameters
With the exception of the amplitude and power, the operating parameters can only be set while in
stand-by mode (idle mode). In order to set or edit the parameters, the desired
fi
eld can be selected
by pressing the "SET" key. The respective function can then be selected using the arrow keys
. If the parameter has number values that can be edited, the editing mode is then activated with
the arrow key . In the marked area (inverse view) of the parameter, the desired value can be