Step 3:
Turn on the homogeniser
Turn on the homogeniser with the power switch.
All saved data and settings will be loaded. Check the settings and load a di
ff
erent program if
necessary, see chapter 3.3.
Step 4:
Activate the ultrasound
Select a small amplitude at the start to prevent any splashing. Observe the maximum allowed
amplitude, see chapter 4.3.
Activate the ultrasound
Warning, risk of splashing!
This is especially the case with small sample quantities and when immersing oscillating probes.
Step 5:
Remove the sample
After sonication, the probe must be removed from the medium. Leaving them in the sonicating
medium for a longer time can cause damage to the probe.
Once the program or time setting has elapsed, the delivery of ultrasound ends automatically.
If continuous sonication has been set, the ultrasound delivery must be stopped manually.
Remove the probe and the temperature sensor, if present, from the sonicating medium.
Step 6:
Clean the probe
In order to prevent contamination with other sonicating media, probes are to be thoroughly
cleaned after every sonication, see chapter 5.2.
Turn o
ff
the homogeniser with the power switch.
Clean the probe and check the wear on the sound emitting surface at regular intervals, see
chapter 5.1.