36 / 46
51376-010 GB/2020-10
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com
Error no.
Possible cause
Remedial action
020
Unknown command (communication
error), remote control command was not
executed or ignored
Observe the application information “Protocol
and command set for remote control” and check
commands.
021
Incorrect command length
(communication error), remote control
command was not executed or ignored
022
Unknown type (e.g. probe)
Check USC, standard/booster horn and probe for
compatibility.
General device errors
Error
Possible cause
Remedial action
Device cannot be switched
on?
(display without function)
No power
Mains cable loose or faulty?
- Check that plug connection is
fi
rmly in
place.
- Check the cable for continuity or,
if needed, exchange the mains
connector.
Device fuse tripped?
- Replace fuses. The fuses are located
in the mains socket on the rear side of
the generator, see chapter 1.
(2 fuses: F2A)
Main fuse faulty?
- Replace main fuse.
Little or no ultrasonic power? Is the connection from the
ultrasonic converter to the
standard/booster horn, or from the
horn to the probe not secure?
- Using the tool supplied, separate parts
from one another, clean the surfaces,
and
fi
rmly screw together once again,
see chapters 4.3.1 ... 4.4.
Standard/booster horn or probe
faulty?
- Check the horn, probe and threaded
pins for cracks, remove and replace
if necessary.
Ensure that the
surfaces are clean and smooth, see
chapter 4.3.1.
Check for erosion on the titanium
fl
at tip / probe - see chapter 5.1.
slight?
- Polish the titanium
fl
at tip and/or
probe.
some pitting?
- Grind or mill the titanium
fl
at tip and/or
probe until
fl
at (max. 1 mm).
heavy?
- Replace
the
titanium
fl
at tip and/or
probe with new ones.
Has liquid seeped in between the
standard/booster horn and the
probe?
- Dismount the probe, clean the
mounting surfaces and threads, dry
and check for evenness, remount the
probe and tighten, see chapter 4.3.1.
Is the threaded pin attachment on
the titanium
fl
at tip defective?
-
Replace the titanium
fl
at tip, see
chapter 4.3.1.
Is the threaded bolt on the
standard/booster horn cracked?
- Disassemble the parts, check the
threaded bolt, replace if necessary,
reassemble the parts, see chapter
4.3.1.
Wrong resonance frequency?
- Conduct frequency scan, see chapter
6.2 (search frequency).
Summary of Contents for Sonoplus HD4050