5 Operation

5.1 Immersion depth notes

The recommended immersion depth for probes is 10 ... 20 mm, to prevent the 

aspiration and mixing-in of air. If mixed-in air is desired, the probe may be 

immersed just a few millimetres.
When producing emulsions, the probe should be immersed to the height of the 

surface separating the liquids to be mixed.

5.2 Preselect parameters for sonication

5.2.1 Set amplitude or power

Information

In the case of greater immersion depths or the sonication of liquids with high vis-

cosities, the probe is more heavily damped. As a result, the preset amplitude may 

not be reached.

The device can regulate the emitted ultrasonic power so that either a set ampli-

tude is kept constant or a set power is kept constant. The amplitude is given in %. 

The absolute value of the amplitude depends on the combination of standard 

horn or booster horn and probe. The power is specified in watts and depends on 

the respective ultrasonic converter. 

1. On the startup screen, tap "Parameters".

2. Tap "Ampl./Power".

3. Tap "Amplitude" if sonication is to be carried out with a constant amplitude, or 

"Power" if sonication is to be carried out with a constant power. 

4. Select the desired value for amplitude or power. You can move the bar of the 

selected setting parameter or enter a numerical value.

5. Confirm your entry with the "Confirm/Save" button.
With amplitude regulation, the amplitude is kept constant. The power can vary 

depending on the condition of the medium. When the respective maximum power 

is reached, the amplitude is reduced. 

Power control keeps the power constant. The power can vary depending on the 

condition of the medium. When the respective maximum amplitude is reached, 

the power is reduced.

Summary of Contents for SONOPULS HD 5020

Page 1: ...00 en 2022 07 Instructions for use SONOPULS Ultrasonic homogeniser Valid for SONOPULS HD 5020 SONOPULS HD 5050 SONOPULS HD 5100 SONOPULS HD 5200 SONOPULS HD 5400 Dokument 51846 000 Sprache en Version...

Page 2: ...2 36 51846 000 en 2022 07 2022 1 2 GmbH Co KG Heinrichstr 3 4 12207 Berlin Ger many Phone 49 30 76880 0 Fax 49 30 7734699 info bandelin com Certified in accordance with ISO 9001 and ISO 13485...

Page 3: ...nce of wireless communication 8 3 Design and function 9 3 1 Overview 9 3 2 Symbols and buttons on the touch screen 10 3 3 Function 10 4 Preparation for operation 11 4 1 Set up homogeniser 11 4 2 Mount...

Page 4: ...5 6 3 Set the language 21 5 6 4 Setting the time and date 21 5 6 5 Set sleep mode 22 5 6 6 Configuring the network 22 5 6 7 Set temperature unit 22 5 7 Troubleshooting a malfunction 23 6 Maintenance 2...

Page 5: ...ese instructions for use before using the ultrasound bath Pay special attention to chapter 2 Safety If you pass this ultrasound bath on to others please enclose these instruc tions for use Contact you...

Page 6: ...ours may ignite e g on electrical components resulting in an explosion Do not use combustible solvents in open sonication vessels as the operational safety of the device could be compromised Safe extr...

Page 7: ...sing a body part to ultrasound can lead not just to skin damage but also to internal tissue damage On fingers the periosteum can become damaged Do not touch the probe the booster horn or the standard...

Page 8: ...e the probes with care Make sure that the probe does not touch the sonication vessel This can damage the probe and the vessel Foreign bodies on the contact surface between the probe and standard horn...

Page 9: ...f the device 1 Ultrasound generator 2 Touchscreen 3 Connector for ultrasonic converter 4 LAN connection 5 USB Type B port 6 USB Type A port 7 Mains switch 8 Cold device socket with fuse holder 9 Ultra...

Page 10: ...e power or amplitude can be selected as the reference variable If the amplitude is selected as the reference variable it is kept constant by the ultrasonic converter regardless of the load provided th...

Page 11: ...illating system in the housing of the ultrasonic converter twists the ultrasonic converter will be damaged Always use the sickle spanner to secure the standard horn or booster horn Never hold the ultr...

Page 12: ...the sickle spanner to secure the standard horn or booster horn Never hold the ultrasonic converter by the housing when tightening or loos ening the probe Requirement The ultrasonic converter is discon...

Page 13: ...asonic converter 4 Place a sickle spanner on the ultrasonic converter Place the ultrasonic con verter on a fixed tabletop and support the sickle spanner on the tabletop 5 Place a sickle spanner on the...

Page 14: ...he entries with the Confirm Save button Exit the Time Date screen by pressing the Confirm Save button The startup screen appears If the language and time are set the startup screen will appear immedia...

Page 15: ...nstant The amplitude is given in The absolute value of the amplitude depends on the combination of standard horn or booster horn and probe The power is specified in watts and depends on the respective...

Page 16: ...hed on and off 5 Confirm your entry with the Confirm Save button 5 2 4 Setting temperature monitoring If you use a temperature sensor you can monitor the temperature of the sonicat ing medium during s...

Page 17: ...on or end it with the Stop button After the end of sonication a screen with a summary of the sonication appears 4 Tap the Confirm Save button The sonication screen reappears with the last settings 5 4...

Page 18: ...rtup screen tap Programs The Programs overview screen is shown 2 Tap on a program number that is highlighted in colour The screen with the summary of program parameters appears 3 Tap the Deactivate bu...

Page 19: ...can still be changed by tapping one of the buttons Amplitude Power Time Pul sation or Temperature Confirm the new settings by tapping the Con firm Save button 3 Optional Tap one of the buttons Amplit...

Page 20: ...data specified under Institution is stored in the documentation issued 1 On the startup screen tap Settings 2 Tap Documentation 3 Tap Institution 4 Enter the desired data 5 Confirm your entry with the...

Page 21: ...3 Tap Brightness 4 Set the desired brightness 5 Confirm your entry with the Confirm Save button 5 6 3 Set the language 1 On the startup screen tap Settings 2 Tap on System 3 Tap Language 4 Select the...

Page 22: ...m Save button 5 6 6 Configuring the network 1 On the startup screen tap Settings 2 Tap on System 3 Tap Interface 4 Tap on Network 5 Activate DHCP if you want the device to automatically obtain the net...

Page 23: ...place if necessary see 4 3 Mounting removing standard horn or booster horn and 4 2 Mounting removing the probe Erosion on the probe If there is minor erosion repolish the sound emit ting surface of th...

Page 24: ...ce This leads to a reduction in sonication efficiency The sound emitting surfaces can be carefully polished or milled a few times You can process the surface using a grinder with a fine grinding disc...

Page 25: ...Before sending the device back to us for inspection repair the device and acces sories must be cleaned pursuant to current laws and regulations and if neces sary must also be disinfected with a surfa...

Page 26: ...waste if it can no longer be used Do not dispose of the ultrasonic bath with household waste Observe the locally applicable regulations for the disposal of electronic waste The TRISON Base contains a...

Page 27: ...0 01 99 59 59 Ultrasound operating modes Pulsating continuous Pulsation time 0 5 600 s Regulation Amplitude or power Amplitude setting range 10 100 in 1 steps Power setting power display Watt Energy d...

Page 28: ...imensions 50 mm 150 mm Degree of protection IP 30 Ultrasonic frequency 20 kHz Weight 0 6 kg Dimensions 50 mm 185 mm Degree of protection IP 30 Ultrasonic frequency 20 kHz Weight 1 0 kg Dimensions 70 m...

Page 29: ...MC Guideline 2011 65 EU RoHS Directive A declaration of conformity can be requested from the manufacturer by providing the serial number Overvoltage category II Degree of contamination 2 Permissible a...

Page 30: ...amplitude mSS MS 1 5 1 5 0 1 10 65 MS 2 0 2 0 0 25 20 70 MS 2 5 2 5 0 5 25 72 Probe sound emitting surface mm Sonication volume ml Max amplitude mSS TS 102 2 0 5 20 125 TS 103 3 1 25 118 TS 104 4 5 3...

Page 31: ...ude mSS TS 103 3 5 90 280 TS 104 4 5 5 100 235 TS 106 6 10 350 220 TS 109 9 10 500 200 TS 113 13 20 900 140 TT 213 13 20 900 105 TS 216 16 25 900 90 TS 219 19 25 900 79 TS 225 25 30 1000 60 Probe soun...

Page 32: ...ies please refer to the product information appendix to these Instructions for use Stand HG 40 for correct variable positioning of the ultrasonic converter Sound proof box LS 40 reduces noise emission...

Page 33: ...during mounting disassembling of the probes Sickle spanner HS 40 42 For holding SH 100 G SH 200 G in place during mounting disassembling of the probes Sickle spanner HS 40 42 long 2 pieces Mounting d...

Page 34: ...9 Accessories 34 36 51846 000 en 2022 07...

Page 35: ...51846 000 en 2022 07 35 36 Accessories 9...

Page 36: ...51846 000 en 2022 07 1 2 GmbH Co KG Heinrichstr 3 4 12207 Berlin Germany Phone 49 30 76880 0 Fax 49 30 7734699 info bandelin com www bandelin com...

Reviews: