23 / 42
51506-002 GB/2020-10
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com
4.3
Selecting a suitable probe
Detailed information on the individual probes can be found in the separate product information.
HD 2070.2
Probe
Ø
MS 72
2 mm
MS 73
3 mm
KE 76
6 mm
TT 13
13 mm
VS 70 T
13 mm
TT 13 FZ
13 mm
BR 30
Connection
to standard
horn
SH 70 G, (FZ 5 G)
FZ 5 G
-
Sonicating
vol.
1 ml -
25 ml
2 ml -
50 ml
5 ml -
100 ml
10 ml -
200 ml
10 ml -
200 ml
-
-
Amplitude*
285
μ
m
ss
245
μ
m
ss
191
μ
m
ss
93
μ
m
ss
97
μ
m
ss
93
μ
m
ss
10
μ
m
ss
Maximum
permitted
amplitude
setting
95 %
100 %
100 %
100 %
100 %
100 %
100 %
HD 2200.2
Probe
Ø
MS 72
2 mm
MS 73
3 mm
KE 76
6 mm
TT 13
13 mm
VS 70 T
13 mm
TT 13 FZ
13 mm
VS 190 T
19 mm
TT 25
25 mm
VS 200 T
25 mm
Connection
to standard
horn
SH 213, SH 213 G, (FZ 7 G)
FZ 7 G
SH 219 G
SH 225 G
Sonicating
vol.
2 ml -
30 ml
5 ml -
90 ml
10 ml -
350 ml
20 ml -
900 ml
20 ml -
900 ml
-
25 ml -
900 ml
30 ml -
1000 ml
30 ml -
1000 ml
Amplitude*
286
μ
m
ss
308
μ
m
ss
255
μ
m
ss
165
μ
m
ss
170
μ
m
ss
165
μ
m
ss
81
μ
m
ss
53
μ
m
ss
51
μ
m
ss
Maximum
permitted
amplitude
setting
50 %
65 %
75 %
100 %
100 %
100 %
100 %
100 %
100 %
HD 2200.2
Probe
Ø
BR 30
BB 6
Connection
to standard
horn
-
-
Sonicating
vol.
-
-
Amplitude*
10
μ
m
ss
7
μ
m
ss
Maximum
permitted
amplitude
setting
100 %
100 %
*
Amplitude values may di
ff