Amplitude: 20 mm
(Standard when
delivered)
Loosen screw
Amplitude: 12 mm
Setting the amplitude
1
2
3
51336-005 en/2022-03
15 / 25
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com
2.3
Changing the amplitude setting
The amplitude of the roller carriage during shaking is set to 20 mm when the device is
delivered.
Alternatively, an amplitude of 12 mm can also be set.
Disconnect device from mains power.
Unscrew the cover of the shaking
device.
ATTENTION: PE connection.
The amplitude can be changed by
adjusting the position of the screw, see
adjacent image.
Then secure the cover and reconnect the
device with the mains.
2.4
Start-up
Release the coupling on the roller carriage: Grasp the lever 1 behind the bearing cup
2 and pull it upwards. Leave or place the clip bearing 3 back on the ball stud. The
roller carriage can now be taken o
ff
the ultrasonic bath.
Equip the roller carriage with the holding clamps (see Section 4.2) or replace the
clamps if already mounted.
Position the roller carriage back on the rim of the bath and check it for free
movement.
Press/click the bearing cup 2 onto the lever 1 on the ball stud (with clip bearing 3).
Connect the device up to the power supply (grounded socket).
Conduct a function test – turn the device on brie
fl
y. The shaking device must move
evenly and must not wobble or grind on the ultrasonic bath. Then switch device o
ff
again.
If necessary, equip the roller carriage with holding clamps, see Section 4.1.