Protected against access by
instruments to dangerous
components,
protected against solid
foreign bodies with a
diameter of 2.5 mm or larger
Protected from splashes
up to 60° from its vertical
axis
8 / 32
51326c GB/2017-06
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com
1.4
Technical data
SONOREX ultrasonic baths are interference-free and
!
.
Safety: EN 61010-1,
EMC: EN 61326-1
Mains supply:
230 V~ (± 10 %) 50/60 Hz, (115 V upon request), mains cable length 2 m
Protection class:
Class I
Frequency 35
kHz
Oscillating tank:
Stainless steel
Serial number (SN):
See type plate on the rear side of the ultrasonic bath
Degree of protection:
IP 33 according to DIN 60529
Ambient conditions according to EN 61 010-1
Overvoltage category:
II
Degree of contamination:
2
Permissible ambient temperature:
5 to 40 °C
Permissible relative humidity up to 31°C: 80%
Permissible relative humidity up to 40°C: 50%
No dewing.
Only for indoor operation.
Speci
fi
cations for use as a medical device
Name:
Ultrasonic bath SONOREX DIGITEC ... F
UMDNS nomenclature (ECRI / DIMDI):
14-263
Purpose:
See Section 1.2.
Classi
fi
cation (in acc. with
Directive 98/79/EC for IVD):
Device category 5 device
Type, model, serial number, year of manufacture:
See type plate on the rear side of the device
The ultrasonic bath has been inspected pursuant to the currently applicable standards and is to be
installed and put into operation pursuant to EMC directions, see Section 1.4.1.
Speci
fi
cations pursuant to the Medical Devices Operator Ordinance (MPBetreibV):
Startup at location, functional check
and personnel training (§ 5):
Not required
Technical safety controls, (STK, § 6):
No speci
fi
cations
Technical measurement controls, (MTK, § 11): Not applicable