17
4.2.3
Setting the runtimes
The duration of ultrasound emission is defi ned in the
"Time setting" area.
To do so, enter the desired runtime using the slider
control or by entering it numerically in the input fi eld,
and confi rm each time with the Enter key.
To start the ultrasonic bath, a runtime of at least 1
sec. must be set.
The runtime can only be set when the "Continuous
operation" mode is not activated!
Note:
On the ultrasonic bath, the next highest time is always
displayed as a chase light for the remaining time. For
example, if 5 min 1 sec. are entered with WINSONIC
DT, they are displayed on the ultrasonic bath as 10
min. The runtime for the ultrasonic bath matches that of the program, however.
4.2.4
Monitoring of the status
Current operating statuses, notes, and error
messages are issued/displayed in the "Status"
area. This makes monitoring of the ultrasonic bath
possible.
There are three types of status messages:
- Operating
statuses
=
green
Example: "
Ultrasound ON
"
- Notes
=
blue
or
black
Example: "
Connection
established…
"
- Error
messages =
red
Example:
"
Communication error
"
An overview of the possible messages can be
found in the following table.
New messages are inserted at the end of the text, in a new row. The previous texts shift
upward. When there are more than 20 rows, it is possible to navigate through the text
with a scroll bar (appears on the right side of the display).
Input fi elds
Slider control