1676-002 en/2022-05
27/52
Operation
/5
5.10 Empty the oscillating tank
WARNING
Danger of electric shock
– Disconnect the mains plug before lifting the ultrasound bath.
– Do not place the ultrasound bath in a sink for emptying.
– Make sure that no liquid can get into the housing.
CAUTION
Hot sonication fluid and oscillating tank
There is a risk of burns when lifting the ultrasound bath to empty it.
– Allow the ultrasound bath to cool down before lifting it.
Contamination on the bottom of the oscillating tank reduces the
ultrasonic output. Empty and clean the oscillating tank if the soni-
cation fluid is visibly contaminated.
Also, observe the specifications of the manufacturer of the prepa-
ration regarding the service life of the sonication fluid.
Replace used sonication fluid completely. Do not freshen the fluid
by topping it up.
Procedure
1. Switch off the ultrasound. If present, switch off the heating. If
you need to move the ultrasound bath to empty it, disconnect
the mains plug.
2. If the ultrasound bath has an outlet, open the ball valve. If the
ultrasound bath has no outlet, carefully pour out the oscillating
tank contents over the corner.
3. Rinse the oscillating tank thoroughly.
4. Wipe the ultrasound bath dry with a soft cloth.
5. If necessary, disinfect the ultrasound bath with a suitable sur-
face disinfectant.