4.4.2
Separating the oscillating tub from the housing
4.4.2.1 Disassembly of screwed drain
-
Remove the base plate, see item 3.
-
Disassemble the drain: Cut the silicon rubber seal between the edge of the tub and the housing
with a sharp knife and push out the tub.
-
Pull the thermostatic sensor out of the push-in link or remove the regulating thermostat completely,
see item 4. 3. 6
Before disassembling, heat the
locking nuts 3, 10, ball cock 11, nuts
5, 9 and 10 as well as the bended
pipe 2 with a hot air pistol. Then:
-
Remove locking nut 10 from ball
valve
11.
-
Remove ball cock 11.
- Loosen
nut
9 and fix with lock
nut
10.
- Loosen
nut
5 and screw it in
the direction to the bended
pipe 2.
-
Loosen locking nut 3 from bended pipe 2.
-
Unscrew the threaded pipe 4 using spanners for locking nuts 9 and 10 then, place spanner on
locking
nut
9.
- Remove
nuts
3 and 5 so that the threaded pipe can be removed.
-
Unscrew the heated bended pipe 2 with a pipe wrench from the screw neck on the tub 1.
The assembly of the drain is described in 4. 3. 7.
4.4.2.2 Disassembly of welded drain
-
Remove the base plate, see 3.
-
Disassemble the drain: Cut the silicon rubber seal with a sharp knife between the edge of the tub
and the housing and push out the tub.
-
Pull the thermostatic sensor out of the push-in link or remove the regulating thermostat completely,
see 4. 3. 6
Then:
-
Unscrew the ball cock 4 and hold
the locking nut 3 with a spanner.
-
Unscrew the locking nut 3
-
Take out the tub together with
drain (2 = sealing ring).
The case rivets at the drain side must be drilled out
1
2
3
4
and be replaced later when removing the tub from
following cleaners:
RK 158 S, RK 170 /H, RK 510 /H, RK 512 H /CH, RK 513,
RK 514 /H /BH, RK 515 CH RK 517 H, RK 1028 /H /BH /CH, RK 1029 /H, RK 1040, RK 1050 /BH /CH.
The assembly of the drain is described in 4. 3. 7.