4/
Preparation for operation
18/52
1676-002 en/2022-05
3. Briefly switch on the ultrasound. To do this, turn the knob to the
right for the ultrasound duration, and turn it back to “0” after 1
to 2 seconds.
Results
» A distinct sound can be heard when the ultrasound is turned on.
If there is no sound, contact Service.
A foil test should be carried out before first use. This serves to doc-
ument the effect of the ultrasound. See chapter
.
4.4 Rinsing out the oscillating tank
Thoroughly rinse the ultrasound bath’s oscillating tank with water
before first use.
In order to protect the surface during transport and storage, the
ultrasound bath is covered with an oily preservative. Before the
ultrasound bath is put into service, this preservative must be
removed with a suitable cleaning agent.