5/
Operation
26/52
1676-002 en/2022-05
only works where the fluid is in contact with the goods to be
treated or the inset beaker.
• Position the side with greater soiling so it is facing downwards.
Insert objects with joints (e.g. scissors or tongs) in open condi-
tion so that the whole surface is optimally reached by the soni-
cation fluid.
5.9 Remove treated objects
WARNING
Risk of burns
The sonication fluid, objects to be treated, the surface of the ultra-
sound bath and accessories may be very hot.
– Do not touch the surface of the ultrasound bath or accessories
such as the lid. Do not reach into the sonication fluid.
– Allow the sonication objects to cool before touching them.
Switch off the ultrasound before removing the sonication objects.
Do not remove sonication objects by hand. Carefully remove e.g.
the insert basket with the sonication objects and place it on a level
surface.
Rinse sonication objects with clean water.
Do not leave sonication objects in the sonication fluid for too long.
This can damage the objects.