1676-002 en/2022-05
33/52
Maintenance
/6
tions or holes are produced. With sturdier foils (thicker or coated
ones), the sonication time may be up to 3 minutes.
6. Switch off the ultrasound. Remove the foil test frame. Remove
the aluminium foil from the foil test frame and allow it to dry.
7. The foil must be perforated, see
image. Else, we recommend
having the device checked by
the Service department at BAN-
DELIN electronic
GmbH & Co. KG: See chapter
8. Archive the foil with the test
date and serial number of the
ultrasonic bath. The foil test document template can also be
completed and archived.
9. Rinse the oscillating tank thoroughly to remove any detached
foil particles.
You can order suitable foil test frames from BANDELIN electronic
GmbH & Co. KG. The foil test frames are configured for a wide
range of tank dimensions. Aluminium foil is additionally required to
perform the test as it is not included in the delivery.
Type
Order no.
For
FT 1
3190
RK 31/H, RK 52/H
FT 4
3074
RK 100/H, RK 102 H, RK 103, RK 106,
RK 255/H
FT 6
3222
RK 156/BH
FT 14
3084
RK 510/H, RK 512 H, RK 514/H/BH
FT 40
3094
RK 170 H, RK 1028/H/C/CH, RK 1040
FT 45
3204
RK 1050/CH