Damages
• If damage to the ultrasonic bath is detected, do not connect the ultrasonic bath to the mains.
• In the event of defects, disconnect the power plug immediately.
• Repairs are only to be conducted by authorised skilled personnel or by the manufacturer.
• Defective parts may only be replaced with original SONOREX parts.
2 Preparation
Carefully unpack the ultrasonic bath and accessories and inspect them for completeness
or possible transportation damages. If any damages or defects are found, these are to be
immediately noti
fi
ed in writing to the transportation company and to the supplier.
Before startup, the ultrasonic bath is to be left to stand at its operating location for 2 hours so that
it may adapt to the ambient conditions.
2.1
Scope of delivery
1 Ultrasonic bath, optionally with heating - see delivery note
1 Ball valve (brass, galvanized) with hose, as of type RK 102 H, packaged separately with
sealing tape and assembly instructions
1 Instruction
manual
Additional accessories according to order - see delivery note
2.2
Set-up / assembly
• Place the ultrasonic bath atop a
fi
rm, level and dry surface. In doing so
-
observe the maximum weight of the ultrasonic bath, including
fl
uid.
Net weight see technical data chapter 1.4.
-
do not block the air supply below the ultrasonic bath.
-
guard against moisture and wetness - risk of electric shock.
• In the case of ultrasonic baths with a drain outlet, mount the ball valve, hose socket and hose,
which are included in the delivery, pursuant to the enclosed assembly instructions.
2.3
Start-up
Thoroughly rinse the ultrasonic bath's oscillating tank with water before its
fi
rst use.
Verify that the control buttons are in the "off" position, i.e. the switch indicator above or the
toggle switch is set to "0", then connect the ultrasonic bath to the mains (grounded socket).
Conduct function test - brie
fl
y plug in the ultrasound (maximum of 1 to 2 seconds), a hissing
noise should be heard. Set to "0" once again.
It is recommended that a foil test be conducted as part of quality assurance prior to the
fi
rst
use.
This test is to be saved for later comparison, see appendix for information.
If applicable, hang accessories in the ultrasonic bath and place lid on top.