2/3
DIRECTLY
in the oscillating tank:
- Fill the oscillating tank to 2/3 with cleaning liquid
(TICKOPUR, STAMMOPUR see dosage on label).
Improperly filled units can suffer damage (see point 7 –
dosing table).
Cleaning with insert basket
2/3
- Use
always an insert basket or utensil holder.
- Parts must not be placed on the tank bottom (Exemption
basket K6 and SH 7).
INDIRECTLY
in inset beakers or non-perforated insert plastic tub
⇒
SONOREX
Accessories
- Fill the oscillating tank with contact liquid (for example TICKOPUR R 33 diluted with
water, 1 % solution).
- Fill the inset beakers or insert plastic tub with cleaning liquid (for example TICKOPUR,
STAMMOPUR).
- Suspend inset beakers in the
positioning cover, do not place
them on the tank bottom.
Insert plastic tub
Cleaning
liquid
Contact
liquid
2/3
≥
2cm
2/3
- Inset beakers or insert plastic tub
must be immersed in the contact
liquid to a depth of at least 2 cm.
See to the tank filling to 2/3.
- Several inset beakers with different
cleaning liquids can be sonicated at the
same time.
When inserting the objects to be cleaned, be sure to observe the following:
- Place the dirtier side towards the bottom.
- Do not stack - ultrasonic energy will be absorbed.
- Sensitive pieces may not be allowed to touch one another. Avoid damage by using
plastic insets, plastic insert baskets or the knobbed silicone mats in insert basket.
- Be sure that there are no air
bubbles in hollow spaces (for
example pocket holes).
- Check the temperature when
cleaning heat-sensitive objects.
- 15 -