13
2 Preparation
Carefully unpack the ultrasonic/rinsing bath and accessories and inspect them for
completeness or possible transportation damages. If any damages or defects are
found, these must be immediately noti
fi
ed in writing to the transportation company and
to the supplier.
Before startup, the ultrasonic bath is to be left to stand at its operating location for
2 hours so that it may adapt to the ambient conditions.
2.1
Scope of delivery
1 Ultrasonic/rising bath, see delivery note
1 Accessories kit
1 Instructions for Use manual
Additional accessories according to order - see delivery note
2.2
Set-up / assembly
• Place the ultrasonic/rinsing bath on a
fi
rm, level and dry surface. In doing so
- observe the maximum weight of the tank, including liquids.
For net weight, see technical data, chapters 1.4.1 to 1.4.3.
- do not block the air supply below the tank.
- guard against moisture and wetness - risk of electric shock.
• Operation is not permitted if the legs are not screwed on, this will cause the built-in
fan cooling on the bottom of the device to become inoperable.
Minimum distance to the ground: 30 mm!
The ultrasonic/rinsing bath will become critically damaged without cooling.
Set-up
Fully remove transportation aids (pallets, transport safety devices).
The height-adjustable legs provided should be mounted on the bottom of the tank.
The tank must then be horizontally levelled.
Set up the ultrasonic/rinsing bath in a dry room.