6536-000 en/2022-03
19/54
Preparation for operation
/4
Do not remove the TRISON Lift until the TRISON Twist has been removed. Pull the
foot element on the foot plate forward until the TRISON Lift can be removed.
4.5 Attaching and removing the TRISON Twist
Attaching the TRISON Twist
NOTICE
– The plug of the TRISON Twist is not waterproof. Make sure that it does not
come into contact with the sonication fluid. If the plug has come into contact
with liquid, allow it to dry completely before connecting it. The plug can be
blown out with compressed air.
– Do not twist the plug or insert it forcefully. Any damage could lead to a short
circuit!
Requirements
• The TRISON Lift is mounted.
• There are no instruments on the TRISON Twist.
Procedure
1. Hold the TRISON Twist plug firmly and make sure that it does not come into
contact with the sonication fluid.
2. Grip the handle of the TRISON Twist and push it downward into the TRISON
Lift guide, until the TRISON Twist engages.
3. Connect the plug of the TRISON Twist to the TRISON Base.
Removing the TRISON Twist
Requirements
• All instruments have been removed from the TRISON Twist.
• The hose couplings on the TRISON Base have been removed.
Procedure
1. Disconnect the TRISON Twist plug from the TRISON Base. Hold the plug firmly
and make sure that it does not come into contact with the sonication fluid.
2. Grip the handle of the TRISON Twist and pull it upwards into the TRISON Lift
guide, until the TRISON Twist can be removed.