6536-000 en/2022-03
21/54
Operation
/5
5.1.2 Pour in the sonication fluid
CAUTION
Risk of scalding
– Do not fill the oscillating tank with hot water.
– Maximum filling temperature: 50 °C.
NOTICE
Damage to the ultrasonic bath caused by condensation
At high humidity, condensation forms on the outside of the oscillating tank when
cold water is added.
– Do not fill the oscillating tank with cold water at high humidity.
NOTICE
Damage to the oscillating tank
If using a powder preparation, do not add it directly to the oscillating tank.
– Mix a powder preparation in another container before adding it to the oscillat-
ing tank.
– Add the preparation to the oscillating tank only when it is completely dis-
solved.
Requirements
• The outlet must be closed.
• The ultrasonic bath must be turned off.
Procedure
1. Fill
1
⁄
3
of the oscillating tank with water.
2. Dose the preparation into the oscillating tank. See chapter
3. Fill with water up to the filling level mark; avoid foam formation when doing
so.
5.1.3 Degas the sonication fluid
Freshly-filled sonication fluid or fluid that has remained in the oscillating tank for
a long period of time must be degassed prior to use. Degassing of the sonication
fluid increases the ultrasound effect.
Procedure
1. Cover the oscillating tank with the lid, if available.
2. Select the “Degassing” program on the TRISON Base.
3. Fill the oscillating tank if needed, see chapter
4. Tap “START” to start degassing.