5
/
Operation
26/54
6536-000 en/2022-03
2. Carefully slide the closed MIS instruments into the adapters so that the mov-
able instrument tip is fully visible in the inspection glass. Make sure to insert
the instruments straight so as not to damage the adapter seals.
Fig. 15 Pushing the MIS instrument into the adapter
3. Close the adapter seals by turning the outer rotation ring by three notches in a
clockwise direction.
» You will hear a snapping sound every time you click into a notch.
Fig. 16 Closing the adapter seal
Caution!
Turn the outer rotation ring by only three notches in a clockwise
direction. If the rotation ring is turned too far, the adapter seal may be dam-
aged.
4. Open the instrument tips.
5. Place the TRISON Rack in the oscillating tank so that the instruments are
completely immersed in the sonication fluid.
6. Connect both hose couplings to the TRISON Base. Make sure that the hose
couplings engage correctly.
7. Place the ends of the two return hoses loosely in the sonication fluid.
8. Select the “MIC” cleaning program on the TRISON Base.
9. Select the rinsing channels of the connected instruments on the touchscreen.
10. Tap “START” to start the cleaning program.
» The cleaning process starts. A progress bar is displayed on the touchscreen.
» After the cleaning process has finished, information on all instruments is
displayed.
Check on the display whether the instruments have been rinsed or are
clogged, see below.
11. As soon as cleaning is complete, remove the TRISON Rack from the sonication
fluid.
12. Open the adapter seals by pulling the rotation ring of the adapter three times
and allowing the rotation ring to spring back.
13. Close the instrument tips. Remove the MIS instruments from the adapters.
Rinse the instruments with water to remove sonication fluid residues.
1
2
3