Operation
/5
5.2.4 Cleaning standard instruments
Requirements
• The basket holder is inserted in the oscillating tank.
• The insert basket for standard instruments is ready for use.
• The oscillating tank is filled.
• The sonication fluid is degassed.
Procedure
1. Distribute the instruments in the insert basket. Make sure that the instru-
ments are open and disassembled if necessary. Place the more heavily soiled
side facing down.
2. Place the insert basket on the basket holder in the oscillating tank so that the
instruments are immersed in the sonication fluid.
3. If necessary, connect the temperature sensor to the TRISON Base. Secure the
temperature sensor to a basket mount on the oscillating tank using the fas-
tening clamp.
4. Select the “Standard” cleaning program on the TRISON Base.
5. Select the sonication time on the touchscreen.
6. Tap “START” to start the cleaning program.
» The cleaning process starts. A progress bar is displayed on the touchscreen.
7. As soon as the cleaning is finished, remove the insert basket from the sonica-
tion fluid.
8. Rinse the instruments with water to remove sonication fluid residues.
Instrument is clogged, cleaning must be repeated.
Instrument is unobstructed, cleaning is complete.