5
/
Operation
28/54
6536-000 en/2022-03
5.3 After the ultrasonic cleaning
5.3.1 Empty the oscillating tank
Dirt on the bottom of the oscillating tank reduces the ultrasound output. Empty
and clean the oscillating tank if the sonication fluid is visibly dirty.
Also, observe the specifications of the manufacturer of the ultrasound prepara-
tion regarding the service life of the sonication fluid.
Replace used sonication fluid completely. Do not freshen the fluid by topping it
up.
Procedure
1. Select “Care” and then “Empty” on the TRISON Base touchscreen.
2. Open the outlet.
3. Tap “START” to start emptying.
4. Connect both hose couplings to the TRISON Base and loosely place the rinsing
plugs or the adapters and the ends of the return hoses into the oscillating
tank.
5. Next, rinse the oscillating tank thoroughly with water, see chapter
5.3.2 Rinsing the ultrasonic bath
Information
If an ultrasonic preparation without disinfecting properties was used for cleaning,
you must disinfect the ultrasonic bath and not only rinse it with water, see chap-
Disinfecting the ultrasonic bath
.
Procedure
1. Close the outlet.
2. Place the TRISON Twist or the TRISON Rack in the oscillating tank.
3. Select “Care” and then “Rinse” on the TRISON Base touchscreen.
4. Fill the oscillating tank with water.
5. Connect both hose couplings to the TRISON Base and loosely place the rinsing
plugs or the adapters and the ends of the return hoses into the oscillating
tank.
6. Tap “Continue” to start the rinsing process.
7. Empty the oscillating tank completely after rinsing, see chapter
.