Maintenance
/6
6 Maintenance
6.1 Cleaning and maintenance of the ultrasonic bath
Cleaning the TRISON Base, ultrasound generator and SONOBOARD
• Wipe the surfaces with a damp cloth. Wipe dry with a soft cloth.
• Do not use abrasive cleaning agents, only care products without scouring
additives.
• If necessary, disinfect the surfaces with a suitable surface disinfectant.
Maintaining the oscillating tank
Impurities in the oscillating tank accelerate their wear, can lead to corrosion and
reduce the ultrasound effect. Therefore, please observe the following instruc-
tions:
• Rinse the oscillating tank thoroughly with water after each use. Wipe dry with
a soft cloth.
• Clean edges and remove residues with a stainless steel care product without
abrasive additives.
• Do not use steel wool, scrapers or shavers to clean the oscillating tank.
• Metal parts and rust particles in the oscillating tank cause corrosion. There-
fore, do not leave any metal parts in the oscillating tank. If rust stains are visi-
ble, remove them immediately with a soft cloth and a stainless steel cleaning
product without abrasive additives.
6.2 Tests
NOTICE
Damage to the ultrasonic bath
– Perform tests only on a filled ultrasonic bath.
If one of the tests does not produce the desired result, contact Service. See chap-
ter
.
Checking the ultrasound performance
The power can be checked with a wattmeter between the mains plug of the ultra-
sonic bath and the socket.
Requirement
• The oscillating tank is filled with water.
Procedure
1. Select the “Standard” cleaning program on the TRISON Base. Tap “START” to
start the ultrasound.
2. Read the power.
3. Switch the ultrasound off again.