6
/
Maintenance
36/54
6536-000 en/2022-03
4. Compare the values read with the technical data. See chapter
The measured values may not deviate by more than 20 % from the values in the
Technical Data.
Checking the ultrasound effect
Perform a foil test to check the effect of the ultrasound during commissioning
and at regular intervals. A check every 3 months is recommended. See chapter
Checking the rinsing and moving function
NOTICE
Risk of damage to robotic instruments
– If testing the rinsing and moving function with a robotic instrument, do not
touch the tip of the instrument.
Requirements
• The TRISON Lift and a TRISON Twist are installed.
• The oscillating tank is filled with water.
Procedure
1. If necessary, connect a robotic instrument to the TRISON Twist in order to
better check the moving function.
2. Connect the two hose couplings to the TRISON Base. Make sure that the hose
couplings engage correctly. Place the ends of the return hoses loosely in the
sonication fluid.
3. Connect the plug of the TRISON Twist to the TRISON Base.
4. Select the “Robotics” cleaning program on the TRISON Base.
5. Select the diameter “8 mm” on the touchscreen.
6. Select all rinsing channels by tapping the symbol at the top of the
touchscreen.
7. START the cleaning program and skip the soaking phase by immediately tap-
ping “START”.
8. Check that there is no water leaking from the hoses.
A flow of approximately 350 ml/min should be displayed on the touchscreen.
9. Check that the robotic instrument tip moves.
If you have not connected a robotic instrument, check whether the four driv-
ers on each mount of the TRISON Twist rotate.