6536-000 en/2022-03
7/54
Safety
/2
certain conditions. The specifications of the respective manufacturer provide
information on suitability for ultrasonic cleaning.
• The TRISON ultrasonic bath is not suitable for cleaning or disinfecting contact
lenses.
• Flammable liquids may not be sonicated in the TRISON ultrasonic bath.
• Indirect sonication in the TRISON ultrasonic bath is not permitted.
Possible side effects/restrictions
• Ultrasound does not disinfect. However, processes such as chemical disinfec-
tion may be accelerated in the ultrasonic bath.
• Cavitation erosion can cause surfaces to be mechanically corroded and their
coatings to dissolve.
User group
The TRISON ultrasonic bath is intended for use in health care facilities, e.g. in a
Central Sterile Supply Department (CSSD). It is to be used by trained personnel.
Operating the ultrasonic bath does not pose a risk to pregnant women.
2.2 Reporting obligation in the event of serious incidents
Report serious incidents to BANDELIN electronic GmbH and Co. KG and to the
competent authorities.
2.3 Prevention of cross-contamination and infection
To avoid cross-contamination, regularly clean and disinfect the surfaces of the
ultrasonic bath with a surface disinfectant that has at least bactericidal, yeasti-
cidal and limited virucidal properties. Process accessories such as holders, carri-
ers or baskets in a cleaning and disinfection device.
At higher temperatures, vapours and aerosols that are contaminated with intro-
duced impurities may escape from the ultrasonic bath. This can lead to infections
and disease. Avoid bath temperatures above 40 °C. If necessary, use a lid, suction
device or protective equipment.
2.4 Keep away from children
Children cannot see the dangers related to the ultrasonic bath. Therefore, keep
children away from the ultrasonic bath.