5
/
Operation
22/54
6536-000 en/2022-03
Information
During degassing, the ultrasound noise becomes quieter. This means that the
ultrasound effect increases.
5.1.4 Checking the adapters for MIS instruments
Checking the adapters for MIS instruments
The seals in MIS instrument adapters are subject to wear due to opening, closing
and the effect of ultrasound. Therefore, check the tightness of the adapters
before each cleaning of MIS instruments.
Fig. 11 Adapter testing strips
Requirement
• The oscillating tank is filled.
Procedure
1. Remove the adapters from the TRISON Rack. Check that the adapter seals are
fully open. If an adapter seal is not fully open, pull the rotation ring of the
adapter and allow the rotation ring to spring back. It will rotate slightly toward
the left. Repeat this step until the adapter seal is fully open.
Fig. 12 Adapter seal fully closed, partially closed and fully open