6536-000 en/2022-03
23/54
Operation
/5
2. Insert all test plugs into the adapter openings.
Fig. 13 Inserting the adapter test plug
3. Insert the adapters back into the TRISON Rack.
4. Place the TRISON Rack in the oscillating tank so that the instruments are
completely immersed in the sonication fluid.
5. Connect the two hose couplings to the TRISON Base. Make sure that the hose
couplings engage correctly.
6. On the touchscreen, select the rinsing channels to be tested.
7. Tap “START” to start the adapter check.
Results
» The test status with progress bars is displayed on the touchscreen.
5.2 Cleaning instruments
5.2.1 Cleaning the Si instruments
Requirements
• The TRISON Lift and the TRISON Twist TT 4000 Si are mounted.
• The oscillating tank is filled.
• The sonication fluid is degassed.
Procedure
1. Grip the TRISON Twist by the handle and swivel it upwards.
2. If necessary, open a push handle and place an Si instrument onto the mount
on the TRISON Twist.
Close the push handle to fixate the Si instrument.
3. Insert the rinsing plugs into the Si instruments. Press the plugs firmly into
place.
4. Grasp the TRISON Twist by the handle and swivel it downwards so that the Si
instruments are completely immersed in the sonication fluid.
5. Connect both hose couplings to the TRISON Base. Make sure that the hose
couplings engage correctly.
6. Place the ends of the two return hoses loosely in the sonication fluid.
7. Connect the plug of the TRISON Twist to the TRISON Base.
8. Select the “Robotics” cleaning program on the TRISON Base.
9. Select the smallest diameter of the connected instruments on the touch-
screen.
10. Select the rinsing channels of the connected instruments on the touchscreen.