5
/
Operation
24/54
6536-000 en/2022-03
11. Tap “START” to start the cleaning program.
» Cleaning starts with a soaking phase without ultrasound and without
instrument movement. A progress bar is displayed on the touchscreen.
» After the cleaning process has finished, information on all instruments is
displayed.
Check on the display whether the instruments have been rinsed or are
clogged, see below.
12. As soon as cleaning is complete, grip the TRISON Twist by the handle and
swivel it upwards.
13. Remove the robotic instruments and rinse them with water to remove the
remaining sonication fluid.
5.2.2 Cleaning the Xi instruments
Requirements
• The TRISON Lift and the TRISON Twist TT 4000 Xi are mounted.
• The oscillating tank is filled.
• The sonication fluid is degassed.
Procedure
1. Grip the TRISON Twist by the handle and swivel it upwards.
2. Carefully insert the Xi instruments into the TRISON Twist mounts.
3. Grasp the TRISON Twist by the handle and swivel it downwards without
immersing the Xi instruments in the sonication fluid.
4. Insert the rinsing plugs into the Xi instruments and press them firmly.
5. Grasp the TRISON Twist by the handle and swivel it downwards so that the Xi
instruments are completely immersed in the sonication fluid. When cleaning
the Xi staplers, the TRISON Twist must stand on the feet of the spacer.
6. Connect both hose couplings to the TRISON Base. Make sure that the hose
couplings engage correctly.
7. Place the ends of the two return hoses loosely in the sonication fluid.
8. Connect the plug of the TRISON Twist to the TRISON Base.
9. Select the “Robotics” cleaning program on the TRISON Base.
10. Select the smallest diameter of the connected instruments on the touch-
screen.
11. Select the rinsing channels of the connected instruments on the touchscreen.
12. Tap “START” to start the cleaning program.
» Cleaning starts with a soaking phase without ultrasound and without
instrument movement. A progress bar is displayed on the touchscreen.
» After the cleaning process has finished, information on all instruments is
displayed.
Instrument is clogged, cleaning must be repeated.
Instrument is unobstructed, cleaning is complete.
1
2
3
4
A1
A2
B1
B2
1
2
3
4
C1
C2
D1
D2