Check on the display whether the instruments have been rinsed or are
clogged, see below.
» Stapler instruments are stiffer than other robotic instruments, so the shaft
or distal end of the Xi stapler moves only to a limited extent during cleaning.
13. As soon as cleaning is complete, grip the TRISON Twist by the handle and
swivel it upwards.
14. Remove the robotic instruments and rinse them with water to remove the
remaining sonication fluid.
Information
Xi staplers can also be cleaned on the TRISON Twist TT 4000 Xi. This requires the
Xi spacer which must be ordered separately, see chapter
spacer firmly onto the feet of the TRISON Twist. Due to the larger dimensions,
only one Xi stapler can be cleaned at a time. The Xi stapler must be connected to
one of the middle mounts of the TRISON Twist.
5.2.3 Cleaning rinsable MIS instruments
Requirements
• The adapters have been tested for tightness, see chapter
.
• The oscillating tank is filled.
• The sonication fluid is degassed.
Procedure
1. Check that the adapter seals are fully open. If an adapter seal is not fully open,
pull the rotation ring of the adapter and allow the rotation ring to spring back.
It will rotate slightly toward the left. Repeat this step until the adapter seal is
fully open.
Fig. 14 Adapter seal fully closed, partially closed and fully open
Instrument is clogged, cleaning must be repeated.
Instrument is unobstructed, cleaning is complete.