Operation
/5
5.3.3 Disinfecting the ultrasonic bath
If an ultrasonic preparation without disinfecting properties was used for cleaning,
you must disinfect the ultrasonic bath instead of only rinsing it with water.
Procedure
1. Close the outlet.
2. Place the TRISON Twist or the TRISON Rack in the oscillating tank.
3. Select “Care” and then “Disinfect” on the TRISON Base touchscreen.
4. Fill the oscillating tank with water and a suitable preparation for disinfection.
5. Connect both hose couplings to the TRISON Base and loosely place the rinsing
plugs or adapters and the ends of the return hoses into the oscillating tank.
6. Select a treatment time and tap “START” to start the process.
7. Empty the oscillating tank completely after disinfection, see chapter
8. Rinse the ultrasonic bath with water, see chapter
5.3.4 Cleaning and disinfecting the TRISON Lift, TRISON Twist and
TRISON Rack
Clean and disinfect the TRISON Lift, the TRISON Twist and the TRISON Rack regu-
larly. To do so, you can place them in the oscillating tank when you clean and dis-
infect the oscillating tank, see chapter
and
chapter
Disinfecting the ultrasonic bath
.
5.3.5 Rinsing the filter
The filter is backflushable and can be reused.
After cleaning the robotic instruments or MIS instruments, the filter must be
rinsed daily and checked for damage. If it is damaged, it must be replaced.
Requirements
• The ultrasonic bath is not active.
Procedure
1. Unscrew the transparent filter housing at the bottom on the TRISON Base,
and rinse it with water.
2. Remove the filter.
3. Discard or clean the filter by rinsing the contamination under running water.
4. Insert the new or cleaned filter with the opening facing upwards. Make sure
that it is inserted straight. If the filter is inserted at an angle, it can be dam-
aged.
5. Check that the sealing ring is present in the filter housing, and screw on the
filter housing.