5
/
Operation
30/54
6536-000 en/2022-03
5.3.6 Saving of logs
After each cleaning cycle is completed, a log that summarises important informa-
tion regarding the cleaning process is created. It is stored in the internal memory,
where up to 50 logs can be managed and saved.
The logs can be transferred to a computer using a USB stick or an existing Ether-
net connection.
Retrieving logs via the USB interface
Procedure
1. From the startup screen select “Settings”, then “Users” and then “Documen-
tation”. The number of internally stored logs is shown in the upper left corner
of the touchscreen.
2. Open the USB interface below the touchscreen and insert a USB stick.
» The detected hardware is displayed at the top right of the touchscreen.
3. Tap the grey USB button to transfer the logs to the USB stick.
» The logs are automatically deleted from the internal memory of the TRISON
Base.
4. Once the logs have been transferred, remove the USB stick and close the USB
interface.
Retrieving logs via the Ethernet interface
Requirements
• The Ethernet cable has been plugged in before the TRISON Base is switched
on. Otherwise, the IP address will not be available.
Procedure
1. From the startup screen select “Settings”, then “System”, and then “Informa-
tion” to view the IP address.
» If no IP address is displayed, switch off the ultrasonic bath, check the net-
work connection, and switch the ultrasonic bath back on, see
the ultrasonic bath on and off
2. On a computer, enter the IP address in the browser and confirm with Enter.
» All logs stored internally on the TRISON Base are displayed.
3. Right-click on “view” to save the log under “Save As” on the computer or
other data carrier.
» The logs are not automatically deleted from the internal memory of the
TRISON Base.
4. Click “delete” to delete the logs from the internal memory of the TRISON
Base.