Hydra 2000 User Manual
Part 3 - Calibration
HB-0844-02
3-24
(4) Press
Scroll Down
the lower text now shows CAL
VAL 2 flashing.
(5) Press
Enter
, the lower text now shows OFFSET C.
(6) Press
Enter
, the lower text now shows OFFSET C
flashing and by use of
Scroll Up/Down
the temperature
should be changed to the reference value.
(7) Press
Enter
, the offset value is accepted and the
upper display will show the adjusted measured
temperature.