Overview 

Thank you for selecting the PHscan series pocket pH tester. This user

 

manual provides a step-by-step guide to help you operate the tester, 
please carefully read the following instructions before use. 
 

Installing the Batteries 

1.  Remove the battery compartment lid. 

 

 

2.  Insert three LR44 alkaline batteries into the battery compartment, 
 

note polarity. 

 

 

3.  Replace the battery compartment lid to its original position and
 

turn clockwise until tight. 

 

 

Keypad 

 
Key   

 

Function 

 

  Switch the tester on or off 

  Lock or unlock measurement 

  Exit calibration and return to the pH measurement 

 

  Start calibration 

  Select the pH buffer option 

 

  Confirm the calibration or displayed option 

  Toggle between pH and temperature measurement 

 

modes (For PHscan20 tester only) 

 

 

Prior to Use 

Remove the protective cap (and translucent cover) from the bottom of 
the tester. If some salt crystals deposited on the electrode, rinse with 
tap water to clean these deposits. 

 

 
If tiny air bubbles are present inside the pH-sensitive glass membrane, 
gently shake the tester downward to remove air bubbles. 

 

If the glass membrane has dried out, soak the electrode in 3M KCl or 
pH 4.01 buffer solution for about 30 minutes. If above solutions are 
not available, use the tap water. 
 

 

 

Switching the Tester On and Off 

  Press the 

Meas

 key to switch on the tester. 

  Press and hold the 

Meas

 key to switch off the tester. 

 

 

 

 

If you do not press any key within 8 minutes, the tester will switch off 
automatically to conserve energy. 
 

 

 

pH Calibration 

The PHscan10/20 tester allows 2 points pH calibration. We recommend 
calibrating the tester during the first use or new electrode replaced. For 
better accuracy, regular calibration is recommended. Do not reuse the 
buffer solutions after calibration, contaminants in solution will affect 
the calibration and eventually the accuracy of the measurement. 
 

3M KCl solution 

pH-sensitive 