3   

I   

PHscan10/20 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

  During the measurement, never wipe the pH-sensitive membrane 

 

as this will cause static interference, blot dry with a lint-free tissue 

 

to remove waterdrops on electrode. 

  Press the 

 key, the tester will lock the measurement, the HOLD 

 

icon appears on the display. Press the key again to resume   

 

measurement. 

 

    

  If the display shows ---- indicating the measurement exceeds the 

 

range, remove the tester from the sample immediately. 

 

 

 

Electrode   
Maintenance and Replacement 

 

Cleaning the Electrode 

  Since the pH-sensitive membrane is susceptible to contamination,   

 

make sure that rinse the electrode thoroughly with distilled water 

 

after use. 

  If your sample contains the oil or grease, soak the electrode in mild 

 

detergent or electrode cleaning solution for at least 15 minutes,   

 

then rinse with distilled water. 

  If you do not use the tester for a period longer than 1 month, store 

 

the electrode in 3M KCl solution or electrode storage solution. 

 

 

 

 

DO NOT store the electrode in distilled or deionized water, which 

 

will deplete the hydration layer of the pH-sensitive membrane and 

 

render the electrode useless. 

 

Replacing the Electrode 

If the tester fails to calibrate or gives fluctuating readings, you should 
consider replacing the electrode. 
1.  Twist  the  electrode  collar  counter  clockwise,  pull  the  electrode 
 

away from the tester. 

 

 

  

 
2.  Align the slot on the new electrode, gently push the electrode into 
 

the tester. 

 

 

3.  Twist the electrode collar clockwise until tight. 

 

 

 

 

 

Appendix 

 

Preparation of pH Buffer Solutions 

The PHscan10/20 tester is packaged with pH

 

4.01/7.00/10.01 buffer 

sachets required for calibration. 
1.  Half fill a 250 ml volumetric flask with distilled water and add the 
 

pH 7.00 buffer reagent. 

 

 

2.   Swirl the volumetric flask gently to dissolve the reagent and fill 
 

to the mark with distilled water. 

 

 

 
3.   Cap and upend the volumetric flask several times to mix solution. 
 

 

  Preparation of pH 4.01 and 10.01 buffer solutions are the same as 

 

above.   

  Prepared buffer solution should be stored in hermetically sealed 

 

glass container and avoid direct sunlight. 

 

Preparation of Electrode Storage Solution 

Dissolve 24.6 grams of analytical grade KCl reagent in 100 ml distilled 
water. Add pH 4.01 standard buffer and adjust solution to pH 4. 
 

250 ml 