background image

03

PHscan20

04

PHscan20

4

.

01

7

.

00

10

.

01

Hold

Meas

 

 

Power On/Off

 Press 

 

MEAS 

 

key 

 

to turn on the 

 

tester, the display 

 

shows 

 

measured 

 

values.

 

 

Preparation of pH Standard Buffer Solutions

The tester is packaged with three pH buffer packets required for calibration.

Open the pH7.00 buffer packet, place 

 

the powder 

 

into 

 

 

250ml volumetric

   

 

flask. Pour

 

distilled

 

water

 

250ml

 

to

 

scale

 

line,

 

mix the

 

solution

 

until

 

reagent 

   

 

is completely dissolved.

Preparation

 

of 

pH4.01

 

and

 

10.01

 

standard

 

buffer

 

solutions

 

are

 

the

 

same

 

as

   

 

above. Prepared standard buffer solutions 

 

should be stored 

 

in hermetically

   

 

sealed glass containers.

Press and hold the MEAS key for 5 seconds, the tester will turn off.
If you do not press any key for 8

 

minutes,

 

the tester

 

will

 

automatically turn

   

 

off to conserve energy.

   

 

To disable the auto-off function, please read the SETUP MENU section.

The 

 

tester contains an integrated 

 

system menu that allows you 

 

to customize

each displayed option to meet personal preference.

System Menu

PARAMETER

OPTIONS

pH Buffer

USA Standard (pH4.01, 7.00, 10.01, Default)

NIST Standard (pH4.01, 6.86, 9.18)

Single point calibration

Calibration Point

2 points calibration (Default)

3 points calibration

Do not use this function (Default)

Reset

Restores the parameters back to default settings

Disables the auto-off function

Auto-Off Function

Enables the auto-off function (Default)

Disables the auto-hold function (Default)

Auto-Hold Function

Enables the auto-hold function

Cal

1. Press

 

and

 

hold

 

the

 

CAL

 

key

 

for

 

3

 

seconds

 

to

 

enter

 

system

 

menu,

 

the

 

tester

    shows default pH buffer option.

 Press CAL key again to select the available standard.

2.

Summary of Contents for PHscan20

Page 1: ...come our customers suggestions on how to speed up these efforts Hazardous Substance Statement Warranty The warranty period for PHscan series tester is 12 months from the date of shipment Above warrant...

Page 2: ...ER HOLD DESCRIPTION Indicates tester is in the measurement mode Indicates tester is in the calibration mode Indicates tester is in SETUP mode Indicates the measured value has been frozen Indicates tem...

Page 3: ...or 8 minutes the tester will automatically turn off to conserve energy To disable the auto off function please read the SETUP MENU section The tester contains an integrated system menu that allows you...

Page 4: ...ress ENTER key to confirm the tester shows YES OFF auto off function is enabled points calibration indicating that indicating that the Enter Cal When the auto off function is enabled 8 Press CAL key t...

Page 5: ...e selected 1 point calibration in the system menu 2 Rinse the pH electrode with distilled water press MEAS key to turn on the tester 3 ait for the measured value to stabilize the display shows END Sin...

Page 6: ...the measured reading you can continue to take measurement Electrode Care and Maintenance For best results please always keep the sensor s glass bulb wet Ensure that the sensor is thoroughly washed wi...

Page 7: ...eparation is done ORDER CODE ITEM E PHSCAN ST Standard pH Electrode Used for routine pH measurements e g drinking water rainwater etc E PHSCAN FT Flat Surface pH Electrode Used for measuring the skin...

Reviews: